MEDISPEC ES 100 |
| TEKNİK ÖZELLİKLER | CİHAZIN YAPTIĞI ELİSA TESTLERİ | |
|
|
|
|
SPECIFICATIONS |
|
| PHOTOMETRIC | |
| LINEAR MEASUREMENT RANGE | 0.00 TO 3.0 ABSORBANCE UNITS (A) |
| PHOTOMETRIC ACCUARCY |
+/- (1% OF THE READING + 0.01 A FROM 0 TO 1.5 A) +/- (2% OF THE READING + 0.01 A FROM 1.5 TO 3.0 A) |
| STABILITY | DRIFT OF NO MORE THAN 0.005 A IN 8 HOURS |
| WARM UP TUME | 45 SECONDS, BULIT-IN |
| LIGHT SOURCE | TUNGSTEN LAMP |
| STANDART WAVELENGTHS | MEDISPEC ES 100:405,450,492,630nm |
| TYPE OF FILTERS |
3-CAVITY HARD COAT INTERFAFERENCE FILTER,10nm HALF
BANDPASS TYPICAL |
| VESSEL | SINGLE,DOUBLE OR BREAK APART STRIPS,UP TO 12 WELLS LONG,3 STRIP LOADING CAPACITY |
| SPEED |
READS,CALCULATES AND PRINTS RESULTS FOR UP TO 12 WELLS IN
ABOUT 30 SECONDS |
| DISPLAY | LED,16 SEGMENT ALPHANUMERIC |
| KEYBOARD | 16 KEY,MEMBRANE SWITCH ,4x4 |
| CALCULATION MODES |
SINGLE POINT CALIBRATION,MULTI-POINT CALIBRATION WITH
REGRESSION OR POINT TO POINT CURVE FIT,CUTOFF,% ABSORBANCE MULTI-POINT,UPTAKE |
| FUSE REQUIREMENTS | TWO 1/2A A SLOW BLOW 3AG |
| POWER REQUIREMENTS | 110-120/220-240 VAC,50-60 Hz. 50 W (switch selectable) |
| SERIAL PORT |
OUTPUT ONLY,2400 BAUND,1 START BIT,8 DATA,1 STOP,NO
PARITY NO HANDSHAKE |
| OTHER | |
| SAFETY CERTIFICATION | LISTED TO URL 1262,LABORATORY EQUIPMENT |
| ENCLOSURE | ABS PLASTIC COVER WITH METAL BASE |
| DIMENSIONS | APPROX.9"x12"x3"(23x30x8cm.) 13 lbs (5.9 kg) |
| RECOMMENDED OPERATING TEMP. | 18-35°C |
| RECOMMENDED OPERATING HUMIDITY | LESS THAN 85% |
| ACCESSORIES SOLD SEPARATELY |
DRI-DYE CHECK STRIPS FOR INSTRUMENTS QC, REPLACEMENTS
LAMPS, FUSES THERMAL PAPER,SERIAL CABLE |